System Settings
Table 8-4: System Settings (Continued)
Item
Number of Copies
Burned Informa-
tion on Paper
Default Denoise
Level
Default Sharpen
Level
Language
Show Information
on Viewing Screen
Auto Save Image
Exposure Sound
AutoTrackAP
AutoTrackLT
DAP Unit
8.16.1

About

If you want to know about the product or the workstation software, you can
always use the About button on the top of the System Settings screen.
The About dialog shows the following information:
Table 8-5: The About Dialog
Item
Product Model
XGB6-12-1014, Rev 1.04
Chapter 8 - Workstation Software
Description
Specifies the number of copies to print on the printer
Specifies whether to display the image parameter infor-
mation on the report. Such information includes patient
name, sex, study date time, KV, mA, X-ray plane, etc.
Specifies the default level of noise reduction effect that is
automatically applied to a generated image. The value is
integer, ranging from 0 to 4.
Specifies the default level of sharpening effect that is
automatically applied to a generated image. The value is
integer, ranging from 0 to 4.
Specifies the language of the user interface, including
English, Chinese, Norwegian, and Swedish.
Note:
This setting will take effect after workstation software
is restarted.
Specifies whether to show the image parameters on the
viewing screens. Such information includes patient name,
sex, study date time, KV, mA, X-ray plane, etc.
Specifies whether to automatically save the fluoroscopic
images to the local hard drive right after they have been
generated.
Note:
If this setting is set to "No", you will have to manually
save every image that you want to preserve.
Specifies whether to produce an alarm sound when the
fluoroscopic imaging is initiated.
Specifies whether to enable the auto-track feature on the
AP plane. The auto-track feature automatically detects the
region of interest (ROI) of an image and adjusts the image
parameters to offer better image quality and less dosage
given to the patients.
Specifies whether to enable the auto-track feature on the
LT plane.
Specifies the DAP unit, such as uGy*m^2.
Description