Exposure Lamps and Audible Signals

3.6
Exposure Lamps and Audible Signals
The exposure lamp (yellow lamps on top of the viewing monitors, 1) will il-
luminate when a footswitch pedal (or Exposure key on the control panel) is
pressed.
Continuous Fluoroscopy
In this mode, the exposure lamps constantly light up.
Pulse Fluoroscopy
In this mode, the exposure lamps flash on and off.
There are three types of warning sounds during an exposure. The ringing
patterns of the warning sounds are described below:
When
in normal conditions,
the dose rate exceeds 88 mGy/min,
accumulated fluoroscopy time in a single study
reaches 5 min,
Related Topic: "Fluoroscopy time summary" (s8.1.2, p90)
3.7
USB & Network Ports
The ports that are supported by the G-Arm system are shown in the figure
below:
XGB6-12-1014, Rev 1.04
Figure 3-5: Exposure Lamps
Figure 3-6: USB & Network Port
Chapter 3 - System Components
The G-Arm system sounds
two beeps per second
four beeps per second
eight beeps per second
35