Cleaning; Disinfecting; Error Messages - Kodak 2200 User Manual

Intraoral x-ray system
Hide thumbs Also See for 2200:
Table of Contents

Advertisement

Cleaning

Disinfecting

Error Messages

Error message
COOLING
CAUTION
:
If you turn off power to the system, the microprocessor does not calculate the cooling time, and for
safety reasons considers that the system has not gone into the cooling cycle.
OP. ERROR
plus audible alarm
KV ERROR
POWER ERROR
4-2
Clean the outside of the system with a damp paper towel or soft cloth using
an alcohol-based, non-corrosive cleaner.
If necessary, wipe off surfaces with disinfectant.
CAUTION
:
• Do not allow liquids to drip into the system.
• Do not spray cleaner or disinfectant directly onto the machine.
• Protect the system from contamination using barriers available
from dental distributors.
• Follow the manufacturer's safety recommendations when using the
cleaner or disinfectant.
Table 10. Error messages
Cause
Cooling cycle; this message can appear
during a period of intensive use.
Release of the radiography control
button before the end of the exposure.
The display shows the remaining
exposure time. (Based on this time,
decide whether to develop the film or
make another exposure.)
The generator's high voltage value is
different from the required value.
Over- or under-tension of the main
power.
Power supply problem on the filament.
Cut filament.
SM701_K2200_2_en
How to cancel
Do not turn off the system. The error
message will disappear when the system
returns to a satisfactory temperature.
Press any key on the arch to stop the
alarm.
Turn off the system and restart. If the
problem persists, call a qualified service
technician and discontinue using the
equipment.
Turn off the system and restart. If the
problem persists, call a qualified service
technician and discontinue using the
equipment.