Test Result Record; Figure 4.1 Self-Calibration Menu - Rigol DHO1000 Series Manual

Digital oscilloscope
Hide thumbs Also See for DHO1000 Series:
Table of Contents

Advertisement

- Click or tap Start, and then the oscilloscope will start to execute the self-
calibration program.
- After starting the self-calibration program, you can click or tap Exit to cancel
self-calibration operation at any time.
3. After completing the self-calibration, restart the oscilloscope. In the Horizontal
menu, select "Average" for the Acquisition item. Then click or tap the input field of
the Averages item to set the number of averages to 16 with the pop-up numeric
keypad.
4. Set the vertical scale of each channel to 2 mV/div and view the offset of the
waveform of each channel. If the offset is greater than 0.5 div, check whether there
are interference signals around you and whether the power source is well
grounded. After that, perform self-calibration again.
5. Click or tap Close to close the self-calibration information window.
4.2

Test Result Record

Record and keep the test result of each test. In the last chapter of this manual, a test
result record form is provided. The form lists all the test items and their
corresponding performance limits as well as spaces for users to record the test
results.
TIP
It is recommended that users photocopy the test record form before each test and record the
test results in the copy so that the form can be used repeatedly.
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.

Figure 4.1 Self-calibration Menu

Overview
DHO1000 Performance Verification
9