L3Harris NARDA AMS-8061 User Manual page 95

Area monitor selective
Table of Contents

Advertisement

Very high fields, which frequency is out of the overall frequency range
of the created table (from minimum frequency of first band to the
maximum frequency of the last one) but within the receiver range (100
kHz – 6 GHz) may cause saturation of the receiver causing wrong
measurements.
Please consider the sweep time, and consequently the power
consumption, is directly related to the number and size of the
frequency bands.
Frequencies are expressed in MHz.
Click on a specific band to edit band parameter as follows:
Even though it is not advisable to set overlapping bands, it can be
done in specific situations, such as, for example, when a certain
portion of the frequency spectrum must be deeply investigated.
Process time depends on band configuration.
This parameter limits the Rate selection, as it is, obviously, not allowed
having a number of frequencies that requires a sweep time longer than
about a half of the rate selected.
The minimum span (difference between stop and start frequencies)
allowed is 110 kHz.
It is not advisable to have a band crossing the frequency of 20.0 MHz
since this is the frequency where the internal receiver switches
between low-range module and high-range module.
8061SW-02 - User's guide and Data Display
3-45

Advertisement

Table of Contents
loading

Table of Contents