3.12.4 Rate Settings
Configuration of data memorization.
Data are saved in the station at a frequency (Rate) ranging from 1 to 15
minutes.
The Storing Setting affects the battery duration and memory, before old
data are overwritten.
P.Time: Processing time. It mainly depends on the Bands setting. The
minimum Rate should be, at least, =10 x Processing time.
As shown below, if an Averaging Period of 6min and a rate of 1min are
selected the indication will represent, minute by minute, the field average
calculated on the previous 6 minutes.
The selection of a different Averaging Period as well as a different Rate will
invalidate all the data stored with the previous settings that will be no more
available for downloading, therefore the software will indicate this risk.
The following message appears:
If the answer is YES, the software will show another message asking
whether to continue with the new setting, as follows:
Answering YES again, the software will allow you to change the setting but
all the data previously acquired will no longer be available for new
downloads. It is therefore suggested to download any measurement result
before changing settings.
3-38
8061SW-02 - User's guide and Data Display