L3Harris NARDA AMS-8061 User Manual page 78

Area monitor selective
Table of Contents

Advertisement

3.12 Description of
controls
3-28
8061SW-02 - User's guide and Data Display
The control window is divided in different zones grouping the available
commands by type and function. The main zones are:
ALARMS divided in:
Last Field
WARNING
ALARM
Averaging Period
NOTIFY ALARMs through
Battery Section
Probe Section
Temperature Limit
Relative Humidity Limit
Case OPEN
The software is also enable to notify alarm to mobile phone via SMS
(for further information see § Schedule for SMS on pag, 3-37). For each
alarm a string is sent; for full list see ?STA command on the pag. 5-13.
Rate Settings consisting of the options:
1 minute
2 minutes
6 minutes
15 minutes
The Rate Setting affects use of the memory and battery duration.
P.Time: Processing time. It mainly depends on the Bands setting. The
minimum Rate should be, at least, =10 x Processing time.
Schedule for MODEM divided in:
Telephone number
Time (HH:mm)
Stand by (Q) indicated in number of quarters of an hour; this value
affects battery duration
Every (H)
Originate CALL
Up to ten fixed phone numbers can be programmed. The first phone
number which is free and allows the data download will interrupt the search
for the other number.
If you desire to use this function, it is suggest to set the hour of
programmed switching on in the window "schedule for MODEM" at
least 1 hour and 30 minutes after the monitor station was switched
on.
To prevent excessive battery consumption a limit of 24 quarters of
hour for continuous modem ON condition was established.

Advertisement

Table of Contents
loading

Table of Contents