D07
Byte +9
The 16 bit figure named Peak is shown as follows:
D15
Peak
Byte +10/+11
It is the Peak value detected in the stored interval, of field strength measured in the band. The number
should be considered as a Little-endian unsigned 16 bit integer multiplied by the Scaler above
calculated. In the particular case in which the figure Peak is equal to 0xFFFF then Peak value must
be considered invalid (AMS-8061 was not able to get a measurement) and the data are meaningless.
If, for example, 16 bit figure named Peak is 325 and the Scaler 0.01 then the PEAK field strength
value related to the band will be 3.25.
The 16 bit figure named Avg is shown as follows:
D15
D14
Avg
Byte +12/+13
It is the averaged value, either RMS or AVG, detected in the stored interval, of field strength measured
in the band. The number should be considered as a Little-endian unsigned 16 bit integer multiplied by
the Scaler above calculated. In the particular case in which the figure Avg is equal to 0xFFFF then
Avg value must be considered invalid (AMS-8061 was not able to get a measurement) and the data
are meaningless.
If, for example, 16 bit figure named Avg is 2153 and the Scaler 0.001 then the Avg field strength
value related to the band will be 2.153.
D15 D14 D13 D12 D11 D10 D9
Byte
+14/+15
5-44
SMS Messages and Commands
D06
D05
D14
D13
D12
D11
D13
D12
D11
D10
Reserved
D04
D03
D10
D09
D08
D07
16 bit unsigned integer
D09
D08
D07
D06
16 bit unsigned integer
Reserved
D8 D7
D6
D02
D01
D06
D05
D04
D03
D05
D04
D03
D5
D4
D3
D2
D00
D02
D01
D00
D02
D01
D00
D1
D0