3.13.4 Sub Bands
3-44
8061SW-02 - User's guide and Data Display
The Sub Bands window allows viewing and setting up to 20 frequency
bands to be investigated. When the tab is opened, the actual bands
configuration is shown.
Each band can be activated for the selective analysis by ticking the
corresponding box.
After the band configuration has been downloaded you will see the list of
the bands, each one indicated with its start and stop frequencies.
Click Enable Setting, and enter the proper password, to make any change.
The Refresh button can be used to restore readings.
If any parameter has been modified, pressing Exit the following message
appears:
Press Yes to save changes or No to discard them and restore previous
configuration.
Each frequency band can be selected and edited (pressing the Enable
Setting button) as required, keeping in mind the following few rules:
-
20 frequency bands as maximum;
-
in each band the lower frequency value must be smaller than the
upper frequency one;