1.5 Main Specifications
Frequency range
User-Programmable frequency bands
Sensor type
Sensor dimensions (Ø)
Sensor RF connection
Sensor control
Dynamic range
Measurement range
Sensitivity
Overload
Resolution
Linearity
Frequency response (flatness)
Overall anisotropy (EN50383)
Unit
Out of band attenuation
Rejection
Calibration
Calibration interval
Reading rate
Measuring parameters
EMF stored values
Average
Average time
Storing rate
Max. logging before overwriting
Alarms
Tables 1-1 – 1-2 list the specifications of AMS-8061.
TABLE 1-1 Technical Specifications of AMS-8061
100 kHz – 6 GHz
Up to 20, individual start-stop frequency settings
Triaxial, isotropic antenna system
120 mm
50 Ohm, N-male
Multi-pin connector
> 60 dB in all settings of the attenuator
Depending on the antenna
(please refer to the antenna specifications pag.1-5)
>50 dB, depending on settings
> 20 dB
Performed in EA accredited laboratory: Accredia LAT nr.008
2 years (recommended)
Down to 200 ms (depending on the band setting)
Settable bands and automatic configuration
AVG or RMS, Max value
Arithmetic (AVG) or RMS
1 – 15 minutes
1, 2, 6, 15 minutes
18 months @ 6 minutes storing rate
3 months @ 1 minute storing rate.
SMS and/or data download for: field over limit, open case,
temperature, humidity, low battery, sensor failure.
(in accordance with antenna specifications)