Chapter 3, Operating Overview
Triggering CDMA Measurements In Remote Mode (GPIB Operation)
2. Trigger all active CDMA measurements
GPIB Syntax
"TRIG" !triggers all active CDMA measurements.
GPIB Example Program
10 OUTPUT 714;"DISP CRXT"
20 OUTPUT 714;"TRIG:MODE:RETR SING"
30 OUTPUT 714;"TRIG"
40 END
GPIB Example Program Comments By Line
10 !Selects CMDA CELLULAR MOBILE RECEIVER TEST screen
20 !Sets the Test Set's remote trigger mode to Single retriggering
30 !Triggers all active CDMA measurements
Preventing GPIB Bus Lockup
GPIB bus lockup is a condition where the GPIB bus and the Active Controller handshake
are in a temporary holdoff state while the Active Controller waits to read the measurement
result from the Test Set.
This Test Set has a feature that allows the bus to be released automatically when a
measurement cycle is unsuccessful. It is called Indefinite Measurement and is
automatically ON when the Test Set is powered on and causes the Test Set to send the
number +1.7976931348623157 e+308 for real number measurements, and 2147483647
for integer number measurements.
GPIB Example
The following GPIB command turns the Indefinite Measurement feature ON
"CONFigure:MEASure:INDefinite 'On'"
86