Chapter 6, IBASIC Controller
PROGram Subsystem
:EXECute <program_command>
from an external controller, an IBASIC program command in the Test Set's built-
in IBASIC Controller .
<program_command> is string data representing any legal IBASIC command. If
the string data does not represent a legal IBASIC command, an
-285 Program syntax error
Any IBASIC program in the Test Set must be in either the PAUSed or STOPped
state before the external controller issues the :EXECute <program_command>
command. If the IBASIC program is in the RUN state, an
Program currently running
Syntax
PROGram[:SELected]:EXECute <delimiter><program_command><delimiter>
The following notation is used in the command description:
Example
OUTPUT 714;"PROGram:SELected:EXECute 'CLEAR SCREEN'"
OUTPUT 714;"PROG:EXEC 'CLEAR SCREEN'"
320
S:\agilent\e8285\USRGUIDE\BOOK\CHAPTERS\ibasic.fb
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
<delimiter> = IEEE 488.2 <string data> delimiter, single quote or double quote, must
be the same.
or
The :EXECute command is used to execute,
is generated.
IBASIC Error: -284
is generated.
IBASIC Error: