Error Messages
List of Text Only Error Messages
Operation errors generally occur when you try to do something the Test Set was not
designed to do. Most messages tell you what to do to correct the problem, (turn something
off, reduce a field's value, press a certain key, and so forth).
Some common messages are listed here:
All self tests passed.
This message appears if the Test Set did not detect any hardware or firmware failures
during its initial self-diagnostics. This message should always appear immediately after
instrument turn on.
Cal file checksum incorrect - initializing file.
This message usually appears after the Test Set firmware ROM is changed. It is not a
problem in that instance, but should not re-appear during subsequent operation of the Test
Set.
Cannot change parameter while measurement is armed.
This message appears if an attempt was made to change one of the following CDMA
SWEPT POWER MEASUREMENT screen fields after arming the measurement:
Amplitude, Slope, or RF Channel/RF Anl Freq.
Cannot make call at this time.
This message appears if a call is attempted while the Test Set is performing another call
processing function, such as registration. Press the End call key and attempt the call again.
Cannot perform operation during hard handoff.
This message appears if a signaling operation is initiated while a handoff is in progress.
The initiated operation does not proceed.
Cannot perform operation without mobile registration.
This message appears in the message screen if the Page Send field in the CDMA CELL
SIT CONFIGURATION screen is set to AO-AK, AO-NA, GP, AO-AK/GP, or AO-NA/GP
and a valid registration has not been received from the mobile station after the protocol
has been changed on the Test Set or after the Test Set power has been cycled.