Chapter 4, Status Reporting
GPIB Status Register Groups
Accessing the Communicate Status Register Group Registers
The following sections show the syntax and give programming examples, using
®
the HP
BASIC programming language, for the STATus commands used to
access the Communicate Status Register Group registers.
Reading the Condition Register
Syntax
STATus:COMMunicate:CONDition?
Example
OUTPUT 714;"STAT:COMM:COND?"
ENTER 714;Register_value
Reading the Transition Filters
Syntax
STATus:COMMunicate:PTRansition?
STATus:COMMunicate:NTRansition?
Example
OUTPUT 714;"STAT:COMM:PTR?"
ENTER 714;Register_value
Writing the Transition Filters
Syntax
STATus:COMMunicate:PTRansition <integer>
STATus:COMMunicate:NTRansition <integer>
Example
OUTPUT 714;"STAT:COMM:PTR 256"