Agilent Technologies E8285A User Manual page 194

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

Chapter 4, Status Reporting
GPIB Status Register Groups
The following sections show the syntax and give programming examples, using
®
the HP
BASIC programming language, for the STATus commands used to
access the Questionable Data/Signal Register Group registers.
Reading the Condition Register
Syntax
STATus:QUEStionable:CONDition?
Example
OUTPUT 714;"STAT:QUES:COND?"
ENTER 714;Register_value
Reading the Transition Filters
Syntax
STATus:QUEStionable:PTRansition?
STATus:QUEStionable:NTRansition?
Example
OUTPUT 714;"STAT:QUES:PTR?"
ENTER 714;Register_value
Writing the Transition Filters
Syntax
STATus:QUEStionable:PTRansition <integer>
STATus:QUEStionable:NTRansition <integer>
Example
OUTPUT 714;"STAT:QUES:PTR 256"