Agilent Technologies E8285A User Manual page 183

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

Chapter 4, Status Reporting
GPIB Status Register Groups
Reading the Event Register
Syntax
STATus:OPERation:EVENt?
Example
OUTPUT 714;"STAT:OPER:EVEN?"
ENTER 714;Register_value
Clearing the Event Register
The EVENT register is cleared whenever it is queried or whenever the Common
Command *CLS is sent to the Test Set.
Reading the Enable Register
Syntax
STATus:OPERation:ENABle?
Example
OUTPUT 714;"STAT:OPER:ENAB?"
ENTER 714;Register_value
Writing the Enable Register
Syntax
STATus:OPERation:ENABle <integer>
Example
OUTPUT 714;"STAT:OPER:ENAB 256"
Clearing the Enable Register
The ENABLE register is cleared by writing to it with an integer value of zero.
182
S:\agilent\e8285\USRGUIDE\BOOK\CHAPTERS\hpibstat.fb
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Advertisement

Table of Contents
loading

Table of Contents