Table 4
Calibrating Status Register Group Condition Register Bit Assignments
Is Condition
Bit Number
Register
Implemented?
15
14
13
12
11
10
9
8
7
6
5
4
3
2
a
NO
1
a
NO
0
a
NO
a. Although a Condition register is implemented, the Test Set will not parse out or execute GPIB com-
mands while this operation is taking place. Consequently, a query of the Condition register will not be
executed until after the calibration attempt has completed, and will never indicate a "true" condition.
Always query the Event register, and when the calibration attempt has completed an Event register query
result will be returned and will show which bit was latched by the initiation of a calibration attempt.
.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Condition
Not Used (Always 0)
Unused in the Test Set
Unused in the Test Set
Unused in the Test Set
Unused in the Test Set
Unused in the Test Set
Unused in the Test Set
Unused in the Test Set
Unused in the Test Set
Unused in the Test Set
Unused in the Test Set
Unused in the Test Set
Unused in the Test Set
CDMA Baseband calibration
Channel power calibration
Digital power zeroing
Chapter 4, Status Reporting
GPIB Status Register Groups
Comment
Defined by SCPI Version 1994.0
Calibrating CDMA Baseband Physical
Channels
Calibrating Channel Power
measurement.
Average power measurements zeroed
111