Error −210
Trigger error
Error −211
Trigger ignored
Indicates that a GET, *TRG, or triggering signal was received and recognized by the
device but was ignored because of device timing considerations. For example, the device
was not ready to respond.
Error −212
Arm ignored
Indicates that an arming signal was received and recognized by the device but was
ignored.
Error −213
Init ignored
Indicates that a request for a measurement initiation was ignored as another measurement
was already in progress.
Error −214
Trigger deadlock
Indicates that the trigger source for the initiation of a measurement is set to GET and
subsequent measurement query is received. The measurement cannot be started until a
GET is received, but the GET would cause an INTERRUPTED error.
Error −215
Arm deadlock
Indicates that the arm source for the initiation of a measurement is set to GET and
subsequent measurement query is received. The measurement cannot be started until a
GET is received, but the GET would cause an INTERRUPTED error.
Error −220
Parameter error
Indicates that a program data element related error occurred.
Error −221
Settings conflict
Indicates that a legal program data element was parsed but could not be executed due to
the current device state (see IEEE 488.2, 6.4.5.3 and 11.5.1.1.5).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Error Messages
363