Agilent Technologies E8285A User Manual page 340

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

Chapter 6, IBASIC Controller
The TESTS Subsystem
TESTS Subsystem File Descriptions
Three types of files are used in the TESTS Subsystem to store different types of
information.
Code Files
The first aspect of an automated definition is the code itself. This is just a standard
IBASIC Code file that can reside either on the Memory card, on an external disk
drive connected to the GPIB port of the Test Set, or in an internal RAM disk. The
name of this file is appended with .PGM file extension. This tells the TESTS
Subsystem that this particular file contains program code.
Library Files
A Library indicates all of the available test subroutines in the code, the set of all
parameters that might be entered using the user-interface screens, and all
specifications that might be used by the subroutines in the code to decide if a test
point passes or fails.
Only one Library is defined for each Code file. The name of this file is appended
with a .LIB file extension, telling the TESTS system that this is a Library file.
Also, both the Library and Code file should have the same base name to indicate
the relationship between them.
A Library is required to use the user-interface screen functions of the TESTS
Subsystem. If the program is simple enough that there is no need for user-input, or
if all the user-input is simple enough to be accomplished with INPUT statements,
a [NO LIB] option is available.
Procedure Files
A Procedure allows the user to define which of the test subroutines, parameters,
and specifications defined in the Library will be used to test a specific Radio.
There may be many Procedures defined that use the same IBASIC Code and
Library, each using a different subset of the choices available in the Library. These
files are appended with a .PRC file extension, but are not required to have the
same base name as either the Library or the Code. The name of the corresponding
Library (if any) is stored in each Procedure file.
339
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com