Chapter 4, Status Reporting
GPIB Status Register Groups
The following sections show the syntax and give programming examples, using
®
the HP
BASIC programming language, for the STATus commands used to
access the Calibration Status Register Group registers.
Reading the Condition Register
Syntax
STATus:CALibration:CONDition?
Example
OUTPUT 714;"STAT:CAL:COND?"
ENTER 714;Register_value
Reading the Transition Filters
Syntax
STATus:CALibration:PTRansition?
STATus:CALibration:NTRansition?
Example
OUTPUT 714;"STAT:CAL:PTR?"
ENTER 714;Register_value
Writing the Transition Filters
Syntax
STATus:CALibration:PTRansition <integer>
STATus:CALibration:NTRansition <integer>
Example
OUTPUT 714;"STAT:CAL:PTR 256"