Agilent Technologies E8285A User Manual page 360

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

Error −113
Undefined header
The header is syntactically correct, but it is undefined for this specific device. For
example, *XYZ is not defined for any device.
Error −114
Header suffix out of range
Indicates that a nonheader character has been encountered in what the parser expects is a
header element.
Error −120
Numeric data error
This error, as well as errors −121 through −128, are generated when parsing a data element
which appears to be numeric, including the nondecimal numeric types.
Error −121
Invalid character in number
An invalid character for the data type being parsed was encountered. For example, an
alpha in a decimal numeric or a "9" in octal data.
Error −123
Exponent too large
The magnitude of the exponent was larger than 32000 (see IEEE 488.2, 7.7.2.4.1).
Error −124
Too many digits
The mantissa of a decimal numeric data element contained more than 255 digits excluding
leading zeros (see IEEE 488.2, 7.7.2.4.1).
Error −128
Numeric data not allowed
A legal numeric data element was received, but the device does not accept one in this
position for the header.
Error −130
Suffix error
This error, as well as errors −131 through −138, are generated when parsing a suffix.
Error −131
Invalid suffix
The suffix does not follow the syntax described in IEEE 488.2 7.7.3.2, or the suffix is
inappropriate for this device.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Error Messages
359