Installation Note
Agilent Technologies ESG Vector Signal Generator
Upgrade Kit for Option 002
(Internal Baseband Generator with 32 Msample Memory)
Kit Part Number E4400-60598
Part Number E4400-90550
Printed in USA May 2002
Supersedes April 2002

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies E4400-60598

  • Page 1 Installation Note Agilent Technologies ESG Vector Signal Generator Upgrade Kit for Option 002 (Internal Baseband Generator with 32 Msample Memory) Kit Part Number E4400-60598 Part Number E4400-90550 Printed in USA May 2002 Supersedes April 2002...
  • Page 2 The information contained in this document is subject to change without notice. Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
  • Page 3 To Be Performed By: ..... . (X) Agilent Technologies Service Center (X) Personnel Qualified by Agilent Technologies ( ) Customer Estimated Installation Time: .
  • Page 4: Installation Kit Parts List

    Installation Kit Parts List Item Quantity Description Part Number Installation Note E4400-90550 ESG-C Firmware Upgrade Kit E4400-60597 Option 002 Label 7121-1232 A7 Internal Baseband Generator 32 Msample Memory E4400-60501 W10 Ribbon Cable 8121-0679 For Non Option 1EM signal generators only Cable-BNC W12 8121-0643 Hex Nut 15/21-32...
  • Page 5 Checking Signal Generator Functionality Use the following procedure to confirm that the signal generator powers up and that the internal check identifies no errors. The internal check evaluates the operation of the signal generator and returns an error message if it detects a problem. When signal generators with Option 1E5 are first connected to ac line power, the error NOTE message ERROR 514, Reference Oven Cold occurs which causes both the OVEN COLD...
  • Page 6: Tools Required

    Removing Outer and Inner Covers Tools Required • T-10 driver • T-15 driver • T-20 driver Removing the Outer Cover Refer to Figure 1. Disconnect the power cord. 2. Using a T-20 driver, remove the two strap handles (1) by loosening the screws. 3.
  • Page 7 Figure 1 Outer and Inner Cover Removal...
  • Page 8 Figure 2...
  • Page 9: Installing New Firmware

    NOTE Use the following procedure only if the instrument does not include Option 1EM. Installing the A7 Internal Baseband Generator Board Refer to Figure 2 for this procedure. 1. Locate Top-SLOT 1. 2. Insert the A7 assembly into the front and rear card cage guides. 3.
  • Page 10 Re-Assembling the Instrument Refer to Figure 1. Reinstall the inner and outer instrument covers by reversing the order for removal. 2. Torque all T-10 screws to 9 in-lbs. 3. Torque all T-15 and T-20 screws to 21 in-lbs. 4. Attach the Option 002 label provided to the rear panel, near the original serial number label. Re-Calibrating the Instrument Refer to the ESG Vector Signal Generator Service Guide.