Agilent Technologies E8285A User Manual page 371

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

Error Messages
Error −440
Query UNTERMINATED after indefinite response
Indicates that a query was received in the same program message after a query requesting
an indefinite response was executed (see IEEE 488.2, 6.5.7.5.7).
Error −606
Update of Input Module Relay Switch Count file failed.
Indicates that the Test Set was not able to update the Input Module Relay Switch Count
EEPROM file with the current switch count data from the non-volatile RAM switch count
array. This error is most probably generated as a result of a hardware error or failure. Refer
to the Agilent Technologies E8285A Assembly Level Repair Manual for diagnostic
information.
Error −607
Checksum of Non-Volatile RAM Relay Count data bad.
Indicates that the Test Set was not able to generate the proper checksum for the Input
Module Relay Switch Count data from the non-volatile RAM switch count array. This
error is most probably generated as a result of a hardware error or failure. Refer to the
Agilent Technologies E8285A Assembly Level Repair Manual for diagnostic information.
Error −608
Initialization of Input Module Relay Count file failed.
Indicates that the Test Set was not able to initialize the Input Module Relay Switch Count
EEPROM file during installation of a new input module. This error is most probably
generated as a result of a hardware error or failure. Refer to the Agilent Technologies
E8285A Assembly Level Repair Manual for diagnostic information.
Error −1300
Order attempted while not in Connect state.
Indicates that an attempt was made to send an order type Mobile Station Control Message
(that is - order a change in power level, put the mobile station in maintenance mode, or
send an alert message to the mobile station) when the Call Processing Subsystem was not
in the Connect state.
Error −1301
Handoff attempted while not in Connect state.
Indicates that an attempt was made to handoff a mobile station to a new voice channel
when the Call Processing Subsystem was not in the Connect state.
Error −1302
Release attempted while not in Connect state.
Indicates that an attempt was made to send a Release message to a mobile station when the
Call Processing Subsystem was not in the Connect state.