Error Messages
Invalid channel setting with paging rate set to 'half'.
This message appears if the Sector A Paging field or the Sector B Pilot field is set to
a level between −3.01 and 0 dB with the Page Rate field (on the CDMA CELL SITE
CONFIGURATION screen) set to Half.
Readjust the Sector A Paging field and/or the Sector B Pilot field (on the CDMA
GENERATOR CONTROL screen) so that neither is at a value between −3.01 and 0 dB, or
set the Page Rate field to Full to allow levels up to 0 dB.
Invalid keystroke.
This message appears is a key that has no function relating to the selected field is pressed.
For instance, pressing the Yes On/Off key while the Filter 1 field is selected.
Measurement result is out of range. Adjust Ref Level field to see trace.
This message appears if the Ref Level field of the CDMA SWEPT POWER
MEASUREMENT screen is set to a value that is outside of the display range. (For
instance, the field is set to 0 dB and the level of the trace is more than 15 dB.) This error
message might also appear if no trigger is received and noise is measured.
Mobile reject order received.
This message appears if the mobile station rejected the (extended) handoff direction
message, and the Test Set extinguished the hard handoff indicator and reset the
corresponding status bit to 0. The Test Set Base Station simulation remained in the
original state.
No response from mobile to handoff direction message.
This message appears if the mobile station did not respond to the (extended) handoff
direction message, and the Test Set extinguished the hard handoff indicator and reset the
corresponding status bit to 0. The Test Set Base Station simulation remained in the
original state.
No response from mobile to handoff direction on new channel.
This message appears if the mobile station did not send the handoff completion message,
and the Test Set extinguished the hard handoff indicator and reset the corresponding status
bit to 0. The Test Set Base Station simulation remained in the original state.