Chapter 3, Operating Overview
To Change the Measurement Display
To Specify Units-of-Measure for GPIB Data Transfer
GPIB Units (UNITs command) are used by the Test Set when sending or receiving
numeric values for most field settings and measurement results through
GPIB. Some measurements allow a choice of GPIB Units, but changing GPIB
Units has no affect on the Display Units or Attribute Units settings.
Attribute Units (AUNits command) are used by the Test Set when sending or
receiving numeric values for data functions (Hi and Lo Limits, Reference, Meter,
and Averaging) through GPIB. Some measurements allow a choice of Attribute
Units, but changing Attribute Units has no affect on the Display Units or GPIB
Units settings.
GPIB Example (UNITs command)
The following command changes the GPIB Units for the
measurement to
"DISP CTXT;MEAS:CDM:AVGP:UNIT W"
GPIB Example (AUNits command)
The following command changes the Attribute Units for the
measurement to
"DISP CTXT;MEAS:CDM:AVGP:AUN W"
After receiving this command, the Test Set will use units of Watts for data
functions (such as High and Low Limits).