Error Messages
Error −1310
Timeout occurred while attempting to alert Mobile.
Indicates that the simulated Base Station's internal timer expired before receiving a
response from the mobile station during an alert attempt. The internal timer is set to
20 seconds when the alert order is sent to the mobile station.
Error −1311
RF power loss indicates loss of Voice Channel.
When the CALL CONTROL screen is displayed and the Call Processing
Subsystem is in the Connect state, the host firmware constantly monitors the
mobile station's transmitted carrier power. If the power falls below 0.0005Watts
the simulated Base Station will terminate the call and return to the Active state.
This error message is displayed if the host firmware has detected that the mobile
station's carrier power has fallen below the 0.0005Watts threshold. The call is
dropped and the Call Processing Subsystem returns to the
NOTE:
In order to ensure that the host firmware makes the correct decisions regarding the presence
of the mobile stations's RF carrier, the Test Set's RF power meter should be zeroed when
first entering the Call Processing Subsystem (that is - the first time the CALL CONTROL
screen is selected during a measurement session). Failure to zero the power meter can result
in erroneous RF power measurements. See "Using the Analog Call Processing Subsystem"
chapter in the Agilent Technologies E8285A Application Guide for information on zeroing
the RF Power meter.
Error −1312
Data from RVC contains invalid bits in word [1,2,3].
Indicates that the decoded data received on the reverse voice channel contains invalid bits
in word 1 and/or word 2 and/or word 3. The raw decoded data is displayed in hexadecimal
format in the
data is only displayed when the
Data.
Error −1313
Timeout occurred while in Maintenance state.
Indicates that the simulated Base Station's internal timer expired before the mobile station
was taken out of the maintenance state. The internal timer is set to 20 seconds when the
maintenance order is sent to the mobile station.
Error −1314
Alert attempted while not in Maintenance or Connect state.
Indicates that an attempt was made to send an Alert order to the mobile station when the
Call Processing Subsystem was not in the Maintenance state or Active state.
372
S:\agilent\e8285\USRGUIDE\BOOK\CHAPTERS\errorms.fb
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
top right-hand portion of the
CALL CONTROL
Active state.
screen. Raw decoded
CALL CONTROL
screen
field is set to
Display

Advertisement

Table of Contents
loading

Table of Contents