Chapter 3, Operating Overview
Triggering Analog Measurements In Remote Mode (GPIB Operation)
2. Trigger all active analog measurements (applies if trigger mode is "Single")
GPIB Syntax
"TRIG" !triggers all active analog measurements.
GPIB Example Program
10 OUTPUT 714;"DISP RFAN"
20 OUTPUT 714;"TRIG:MODE:RETR SING;SETT FULL"
30 OUTPUT 714;"TRIG"
40 END
GPIB Example Program Comments By Line
10 !Selects RF Analyzer screen
20 !Sets the Test Set's remote trigger mode to Single retriggering and Full settling
30 !Triggers all active analog measurements
Preventing GPIB Bus Lockup
GPIB bus lockup is a condition where the GPIB bus and the Active Controller handshake
are in a temporary holdoff state while the Active Controller waits to read the measurement
result from the Test Set.
This Test Set has a feature that allows the bus to be released automatically when a
measurement cycle is unsuccessful. It is called Indefinite Measurement and is
automatically ON when the Test Set is powered on and causes the Test Set to send the
number +1.7976931348623157 e+308 for real number measurements, and 2147483647
for integer number measurements.