Chapter 6, IBASIC Controller
PROGram Subsystem
:DEFine?
from the Test Set to an external controller.
The IBASIC program uploaded to the external controller is transferred as IEEE
488.2 Definite Length Arbitrary Block Response Data. The following information
describes some of the characteristics of the IEEE 488.2 Definite Length Arbitrary
Block Response Data type. Refer to the IEEE Standard 488.2-1987 for detailed
information on this data type.
Syntax
PROGram[:SELected]:DEFine?
318
S:\agilent\e8285\USRGUIDE\BOOK\CHAPTERS\ibasic.fb
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The :DEFine? query command is used to upload an IBASIC program
The data starts with a header which begins with a "#", followed by a single non-zero
digit in the range 1-9 which specifies the number of digits in the following count field,
followed by a series of digits in the range of 0-9 which gives the number of data bytes
being sent, followed by the number of data bytes specified by the count field.
Example
#16<data byte><data byte><data byte><data byte><data byte><data byte>
The transfer is terminated by the transmission, from the Test Set to the external control-
ler, of the response message terminator (NL & END messge).
<NL> = new line = ASCII linefeed character.
<END> = IEEE 488.1 END message.