Connecting To The Collision Cell Gas Supply; Required Materials; Connecting The Nitrogen Exhaust Line - Waters Xevo TQD IVD Overview And Maintenance Manual

Table of Contents

Advertisement

6.
Install the 6-mm stud into the regulator outlet.
7.
Connect the free end of the long piece of 6-mm PTFE tubing to the 6-mm stud.
B.5

Connecting to the collision cell gas supply

B.5.1

Required materials

Chemical-resistant, powder-free gloves
Wrench
1/8-inch Swagelok
1/8-inch stainless steel tube (supplied with the instrument)
To connect the collision cell gas supply:
1.
Use the 1/8-inch Swagelok nut and ferrule to connect the 1/8-inch stainless steel tube to the
collision cell gas inlet on the rear of the instrument (see the figure on
2.
Use the wrench to tighten the 1/8-inch Swagelok nut.
3.
Connect the free end of the tube to the collision gas supply.
B.6

Connecting the nitrogen exhaust line

B.6.1
Required materials
Chemical-resistant, powder-free gloves
Sharp knife
10-mm and 12-mm PTFE tubing (included in the Waters Rough Pump Connect Kit)
®
Snoop
(or equivalent) leak detector liquid
®
nut and ferrule
September 7, 2016, 715003560IVD Rev. C
Page 132
page
131).