LCP Status Messages

The following messages may be displayed on the LCP to inform you of the
status of the system. If the LCP is currently in Menu mode, cycle through the
menus to return to Status mode.
No messages will be displayed if an online instrument session is not currently running on
N O T E
the GC/MSD ChemStation.
ChemStation Loading <timestamp>
The Agilent MSD Productivity ChemStation software is starting up.
Executing <type>tune
A tuning procedure is in progress (type = QuickTune or Autotune).
Instrument Available <timestamp>
The Agilent MSD Productivity ChemStation software is not running.
Loading Method <method name>
Method parameters are being sent to the system.
Loading MSD Firmware
The MSD's firmware is being initialized.
The following messages alternately appear on the LCP if the MSD does NOT
complete its bootup sequence properly:
5975T LTM GC/MSD Operation Manual
Server not Found
Check LAN Connection
Seeking Server
Bootp Query xxx
Operating in Electron Impact (EI) Mode