Checkout Chromatograms Fid Checkout Conditions And Chromatogram - Agilent Technologies 6890 Series Service Manual

Gas chromatograph
Hide thumbs Also See for 6890 Series:
Table of Contents

Advertisement

610
Checkout Chromatograms
FID checkout conditions and chromatogram
FID checkout conditions and chromatogram
Column and sample
Type
Sample
Injection volume
Inlet
Temperature
Inlet pressure
Split/Splitless Only
Mode
Purge flow
Purge time
PTV—Splitless Mode
Inlet temperature
Initial time
Rate 1
Final temp 1
Final time 1
Rate 2
Final temp 2
Final time 2
Inlet pressure
Mode
Purge time
Purge flow
Detector
Temperature
H
flow
2
Air flow
Makeup flow, N
Offset
2 of 26
HP-5 30 m × 0.32 mm × 0.25 µm PN 19091J-413
FID Checkout PN 18710-60170
1 µL
250°C
Oven Track
25 psi
Splitless
60 mL/min
0.75 min
40°C
0.1 min
350°C
2 min
250°C
0 min
0.75 min
60 mL/min
300° C
30 mL/min
400 mL/min
25 mL/min
2
Should be < 20 pA
Performance Verification
Agilent 6890 Gas Chromatograph Service Manual
Purged/Packed or Split/Splitless
Cool On-Column
(Constant pressure for EPC inlets,
helium)
720°C/min
100°C/min
25 psi (const. press. for EPC inlets)
Splitless
Jun 2001

Advertisement

Table of Contents
loading

This manual is also suitable for:

6890a6890 plus

Table of Contents