Be prepared to provide the serial numbers of the instruments or devices in
your system when you contact Waters customer support.
To view the information for a system module:
1.
In the Instrument console, select a module from the system tree.
2.
Click Configure > View module information.
Result:
•
Serial number
•
Firmware version
•
Firmware checksum
•
Component software version
Alternatives:
•
From the main window, place the pointer over the visual representation
of the system module for which you want to view information.
•
Obtain the serial number from the printed labels affixed to the rear
panels of system modules or inside their front doors.
Maintenance schedule
Perform the following routine maintenance on the ISM to ensure reliable
operation and accurate results. When using the system throughout the day
(and on nights and weekends), or when using aggressive solvents such as
buffers, perform these maintenance tasks more frequently.
Recommended routine maintenance schedule:
Maintenance
procedure
Replace solvent filters
Clean the air filter in
the door
The Module Information dialog box displays this information:
Required frequency
During scheduled
routine maintenance or
as needed
As needed
For information...
See
page
See
September 26, 2013, 715004208 Rev. A
Maintenance schedule
page 80
and
81.
page
83.
77