Replacing the cartridge in the optional solvent-selection
valve
Warning:
when you handle solvents, change tubing, or operate the isocratic
solvent manager. Consult the Material Safety Data Sheets regarding
the solvents you use.
Caution:
chemical-resistant, powder-free gloves when replacing the
solvent-selection valve cartridge.
Required materials
•
Gloves: clean, powder-free, chemical-resistant
•
Solvent-selection valve cartridge
Required tool
Allen wrench, 2-mm (startup kit)
To replace the cartridge in the optional solvent-selection valve:
1.
In the ACQUITY UPLC Console, select Isocratic Solvent Manager from
the system tree.
2.
Ensure the solvent-selection valve is set to solvent 6.
Warning:
location below the ISM.
3.
Move the solvent bottles to a location below the ISM.
Replacing the cartridge in the optional solvent-selection valve
To prevent injury, always observe Good Laboratory Practice
To prevent contamination to system components, wear clean,
To avoid solvent spills, move the solvent bottles to a
September 26, 2013, 715004208 Rev. A
99
Need help?
Do you have a question about the ACQUITY Isocratic Solvent Manager and is the answer not in the manual?