2 Preparing for Operation
Required materials
•
Cotton swabs
•
Gloves: clean, powder-free, chemical-resistant
•
Nonabrasive, lint-free wipes
To resolve a leak sensor error:
1.
In the Instrument console's Leak Sensors dialog box, confirm that the
ISM's leak sensor detected a leak.
If a leak is detected, a "Leak Detected" error message appears.
Tip:
Caution:
system module, do not disconnect an electrical assembly from the
module while the module remains connected to the ac supply
source.
Follow this procedure to completely interrupt power to the module:
1.
2.
Afterward, wait 10 seconds thereafter before you disconnect an
assembly.
2.
Power-off the ISM.
3.
Open the ISM's door, gently pulling its right-hand edge toward you.
4.
Locate the source of the leak, and make the repairs necessary to stop the
leak.
36
September 26, 2013, 715004208 Rev. A
To avoid damaging electrical components or circuitry of a
Set the module's power switch to Off.
Disconnect the module's electrical supply cord from the ac
source.