Replacing the ISM solvent filter
Warning:
when you handle solvents, change tubing, or operate the ISM. Consult
the Material Safety Data Sheets regarding the solvents you use.
Caution:
handling the solvent filter. Oil from your hands can contaminate the
ISM solvent filter.
Required materials
•
Gloves: clean, powder-free, chemical-resistant
•
New ISM solvent filter
To replace an ISM solvent filter:
1.
Remove the filter end of the solvent tubing from the solvent bottle.
ISM solvent filter:
To prevent injury, always observe Good Laboratory Practices
Wear clean, chemical-resistant, powder-free gloves when
Filter
Replacing the ISM solvent filter
September 26, 2013, 715004208 Rev. A
81