Warning:
the degasser's vent tubing to a fume hood or other suitable
exhaust system, or to a suitable waste container, ensuring the
tubing's discharge end is at all times above the fluid level.
Warning:
vapors, empty the waste container at regular intervals.
Caution:
of waste:
• Place the waste container below the system stack.
• Ensure that the waste and degasser vent tubes are not crimped
or bent. A crimp or bend can impede flow to the waste
container.
• Ensure the exit of the waste and degasser vent tubes is not
immersed in waste solvent. If necessary, shorten each tube so
that no portion of it drops below the top of the waste container
(see next figure).
3.
Route the degasser vent tubing to a fume hood or other suitable exhaust
system, or to a suitable waste container.
Installing the waste tubing and vent tubing for the degasser
To avoid releasing solvent vapors into the room, route
To avoid spills and the subsequent release of solvent
To avoid fluid backup, you must ensure proper drainage