Waters ACQUITY Isocratic Solvent Manager Overview And Maintenance Manual page 26

Table of Contents

Advertisement

1 ACQUITY Isocratic Solvent Manager
3.
The filtered solvent flows through the primary check valve and into the
piston chamber.
4.
The primary piston delivers solvent to the accumulator, the vent valve,
and the inline 22-µL filter during transfer.
5.
The accumulator piston delivers solvent, under pressure, to the vent
valve and the inline 22-µL filter.
6.
From the inline 22-µL filter, the solvent flows to the post-column
addition tee.
7.
The post-column addition tee adds ISM solvent to the solvent.
8.
The diluted solvent then flows into the source of the QDa detector.
26
September 26, 2013, 715004208 Rev. A