Replacing the inline 22-µL filter
Warning:
when you handle solvents, change tubing, or operate the ISM. Consult
the Material Safety Data Sheets regarding the solvents you use.
Caution:
chemical-resistant, powder-free gloves when replacing the 22-μL
mixer/filter.
Required material
Gloves: clean, powder-free, chemical-resistant
Required tools
1/4-inch, open-end wrench
5/8-inch, open-end wrench
22-μL mixer/filter
To replace the 22-μL filter:
1.
Flush the ISM with non-hazardous solvent.
2.
Stop the solvent flow.
3.
Holding the 22-μL filter in place with the 5/8-inch, open-end wrench,
disconnect the inlet compression fitting using the 1/4-inch, open-end
wrench.