Chapter 2 Command System

IEEE488.2 Common Commands

The IEEE 488.2 standard defines some common commands used for querying the basic information of the
instrument or executing the basic operations. These commands usually start with "*" and the keyword of
the command is usually 3-character long.
Command List:

*CLS

*ESE

*ESR?
*IDN?
*OPC
*RST
*SRE
*STB?
*TST?
*WAI
*CLS
Syntax
*CLS
Description
Clear all the event registers in the register set and clear the error queue.
*ESE
Syntax
*ESE <value>
*ESE?
Description
Set or query the enable register for the standard event status register set.
Parameter
Name
<value>
Explanation
The bit 1 and bit 6 of the standard event status register are not used and are always
treated as 0, therefore, the range of <value> are the decimal numbers corresponding to
the binary numbers ranging from 00000000 (0 in decimal) to 11111111 (255 in decimal)
and of which the bit 1 and bit 6 are 0.
Return
The query returns an integer which equals the sum of the weights of all the bits that have
Format
already been set in the register.
Example
*ESE 16
*ESE?
DS1000Z Programming Guide
Type
Range
Integer
0 to 255
/*Enable the bit 4 (16 in decimal) of the standard event status register */
/*The query returns 16*/
RIGOL
Default
0
2-33