RIGOL
:MATH:FFT:HCENter
Syntax
:MATH:FFT:HCENter <cent>
:MATH:FFT:HCENter?
Description
Set or query the center frequency of the FFT operation result, namely the frequency
relative to the horizontal center of the screen. The default unit is Hz.
Parameter
Name
<cent>
Explanation
Current screen sample rate = 100/current horizontal timebase.
Step = horizontal scale of the FFT operation result/50.
Return
The query returns the current center frequency in scientific notation.
Format
Example
:MATH:FFT:HCENter 10000000
:MATH:FFT:HCENter?
Related
:MATH:OPERator
Commands
:MATH:FFT:HSCale
:TIMebase[:MAIN]:SCALe
:MATH:OPTion:STARt
Syntax
:MATH:OPTion:STARt <sta>
:MATH:OPTion:STARt?
Description
Set or query the start point of the waveform math operation.
Parameter
Name
<sta>
Explanation
This command is invalid for the FFT operation. Sending this command will modify the
start points of all the operations (except FFT).
The horizontal axis of the source selected is equally divided into 1200 parts, in which
the leftmost is 0 and the rightmost is 1199.
Return
The query returns an integer.
Format
Example
:MATH:OPTion:STARt 50
Related
:MATH:OPTion:END
Command
2-42
Type
Range
Real
0 to (current screen sample rate×2/5)
/*Set the center frequency to 10MHz*/
/*The query returns 1.000000e+07*/
Type
Range
0 to (End point currently set-1)
Integer
Chapter 2 Command System
Default
5MHz
Default
0
DS1000Z Programming Guide