Rigol DS1000Z Series Programming Manual page 136

Digital oscilloscope
Hide thumbs Also See for DS1000Z Series:
Table of Contents

Advertisement

RIGOL
:TRIGger:RS232:PARity
Syntax
:TRIGger:RS232:PARity <parity>
:TRIGger:RS232:PARity?
Description
Set or query the parity type when the trigger condition is ERRor or PARity in RS232
trigger.
Parameter
Name
<parity>
Explanation
When the trigger condition is PARity, the parity type cannot be set to NONE. At this
point, the default parity type is ODD.
Return Format
The query returns EVEN, ODD or NONE.
Example
:TRIGger:RS232:PARity EVEN
Related
:TRIGger:RS232:WHEN
Command
:TRIGger:RS232:STOP
Syntax
:TRIGger:RS232:STOP <bit>
:TRIGger:RS232:STOP?
Description
Set or query the stop bit when the trigger condition is ERRor in RS232 trigger.
Parameter
Name
<bit>
Return Format
The query returns 1 or 2.
Example
:TRIGger:RS232:STOP 2
Related
:TRIGger:RS232:WHEN
Command
:TRIGger:RS232:DATA
Syntax
:TRIGger:RS232:DATA <data>
:TRIGger:RS232:DATA?
Description
Set or query the data when the trigger condition is DATA in RS232 trigger.
Parameter
Explanation
In 2
Return Format
The query returns an integer.
Example
:TRIGger:RS232:DATA 10
Related
:TRIGger:RS232:WIDTh
Commands
:TRIGger:RS232:WHEN
2-122
Type
Discrete
Type
Discrete
Name
Type
<data>
Integer
n
-1, n is the current data bits and can be 5, 6, 7 or 8.
Range
{EVEN|ODD|NONE}
Range
{1|2}
Range
n
0 to 2
-1
Chapter 2 Command System
Default
NONE
Default
1
Default
90
DS1000Z Programming Guide

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds1104zDs1074zDs1104z-sDs1074z-s

Table of Contents