Rigol DS1000Z Series Programming Manual page 145

Digital oscilloscope
Hide thumbs Also See for DS1000Z Series:
Table of Contents

Advertisement

Chapter 2 Command System
:TRIGger:SPI:TIMeout
Syntax
:TRIGger:SPI:TIMeout <time_value>
:TRIGger:SPI:TIMeout?
Description
Set or query the timeout value when the trigger condition is TIMeout in SPI trigger. The
default unit is s.
Parameter
Name
<time_value>
Return
The query returns the timeout value in scientific notation.
Format
Example
:TRIGger:SPI:TIMeout 0.001
:TRIGger:SPI:TIMeout?
Related
:TRIGger:SPI:WHEN
Command
:TRIGger:SPI:SLOPe
Syntax
:TRIGger:SPI:SLOPe <slope>
:TRIGger:SPI:SLOPe?
Description
Set or query the clock edge in SPI trigger.
Parameter
Name
<slope>
Explanation
POSitive: sample the SDA data on the rising edge of the clock.
NEGative: sample the SDA data on the falling edge of the clock.
Return
The query returns POS or NEG.
Format
Example
:TRIGger:SPI:SLOPe POSitive
:TRIGger:SPI:CLEVel
Syntax
:TRIGger:SPI:CLEVel <level>
:TRIGger:SPI:CLEVel?
Description
Set or query the trigger level of the SCL channel in SPI trigger. The unit is the same as the
current amplitude unit.
Parameter
Name
<level>
Explanation
For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
Return
The query returns the trigger level in scientific notation.
Format
Example
:TRIGger:SPI:CLEVel 0.16
:TRIGger:SPI:CLEVel?
Related
:CHANnel<n>:SCALe
Commands
:CHANnel<n>:OFFSet
DS1000Z Programming Guide
Type
Range
Real
100ns to 1s
/*Set the timeout value to 1ms*/
/*The query returns 1.000000e-03*/
Type
Range
Discrete
{POSitive|NEGative}
/*Set the clock edge to POSitive*/
Type
Range
Real
(± 5 × VerticalScale from the screen center) - OFFSet
/*Set the trigger level to 160mV*/
/*The query returns 1.600000e-01*/
RIGOL
Default
1μs
Default
POSitive
Default
0
2-131