Mask:mdisplay; Mask:sooutput - Rigol DS1000Z Series Programming Manual

Digital oscilloscope
Hide thumbs Also See for DS1000Z Series:
Table of Contents

Advertisement

RIGOL

:MASK:MDISplay

Syntax
:MASK:MDISplay <bool>
:MASK:MDISplay?
Description
Enable or disable the statistic information when the pass/fail test is enabled, or query the
status of the statistic information.
Parameter
Name
<bool>
Explanation
Before executing this command, you need to send the
enable the pass/fail test.
When the statistic information is on, the test results as shown in the figure below will
be displayed at the upper right corner of the screen.
You can sent the :MASK:PASSed?,
query the test results.
Return
The query returns 0 or 1.
Format
Example
:MASK:MDISplay ON
Related
:MASK:ENABle
Commands
:MASK:PASSed?
:MASK:FAILed?
:MASK:TOTal?

:MASK:SOOutput

Syntax
:MASK:SOOutput <bool>
:MASK:SOOutput?
Description
Turn the "Stop on Fail" function on or off, or query the status of the "Stop on Fail"
function.
Parameter
Name
<bool>
Explanation
ON: when failed waveforms are detected, the oscilloscope will stop the test and enter
the "STOP" state. At this point, the results of the test remain the same on the screen
(if the display is turned on) and the [Trigger Out] connector (if enabled) at the rear
panel outputs a single pulse.
OFF: the oscilloscope will continue with the test even though failed waveforms are
detected. The test results on the screen will update continuously and the [Trigger
Out] connector at the rear panel outputs a pulse each time a failed waveform is
detected.
Return
The query returns 0 or 1.
Format
Example
:MASK:SOOutput ON
2-48
Type
Range
Bool
{{0|OFF}|{1|ON}}
Type
Range
Bool
{{0|OFF}|{1|ON}}
:MASK:ENABle
:MASK:FAILed?
and
:MASK:TOTal?
Chapter 2 Command System
Default
0|OFF
command to
commands to
Default
0|OFF
DS1000Z Programming Guide