RIGOL
:CALibrate Commands
Command List:
:CALibrate:QUIT
:CALibrate:STARt
:CALibrate:QUIT
Syntax
:CALibrate:QUIT
Description
Exit the calibration at any time.
Related
:CALibrate:STARt
Command
:CALibrate:STARt
Syntax
:CALibrate:STARt
Description
The oscilloscope starts to execute the self-calibration.
Explanation
The self-calibration can make the oscilloscope quickly reach its optimum working
state to obtain the most accurate measurement values.
During the self-calibration, all the channels of the oscilloscope must be disconnected
from the inputs.
The functions of most of the keys are disabled during the self-calibration. You can
send the
Related
:CALibrate:QUIT
Command
2-8
:CALibrate:QUIT
command to quit the self-calibration.
Chapter 2 Command System
DS1000Z Programming Guide