Rigol DS1000Z Series Programming Manual page 107

Digital oscilloscope
Hide thumbs Also See for DS1000Z Series:
Table of Contents

Advertisement

Chapter 2 Command System
:TRIGger:SLOPe:TUPPer
Syntax
:TRIGger:SLOPe:TUPPer <time>
:TRIGger:SLOPe:TUPPer?
Description
Set or query the upper limit of the time in slope trigger. The default unit is s.
Parameter
Name
<time>
Explanation
This command is available when the trigger condition (refer to :TRIGger:SLOPe:WHEN) is
PGLess and NGLess.
Return
The query returns the upper limit of the time in scientific notation.
Format
Example
:TRIGger:SLOPe:TUPPer 0.000003
:TRIGger:SLOPe:TUPPer?
Related
:TRIGger:SLOPe:WHEN
Commands
:TRIGger:SLOPe:TLOWer
:TRIGger:SLOPe:TLOWer
Syntax
:TRIGger:SLOPe:TLOWer <time>
:TRIGger:SLOPe:TLOWer?
Description
Set or query the lower limit of the time in slope trigger. The default unit is s.
Parameter
Name
<time>
Explanation
This command is available when the trigger condition (refer to :TRIGger:SLOPe:WHEN) is
PGLess and NGLess.
Return
The query returns the lower limit of the time in scientific notation.
Format
Example
:TRIGger:SLOPe:TLOWer 0.000000020
:TRIGger:SLOPe:TUPPer?
Related
:TRIGger:SLOPe:WHEN
Commands
:TRIGger:SLOPe:TUPPer
DS1000Z Programming Guide
Type
Range
Real
16ns to 10s
Type
Range
Real
8ns to 9.99s
/*Set the upper limit of the time to 3μs*/
/*The query returns 3.000000e-06*/
/*Set the lower limit of the time to 20ns*/
/*The query returns 2.000000e-08*/
RIGOL
Default
1μs
Default
992ns
2-93

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds1104zDs1074zDs1104z-sDs1074z-s

Table of Contents