:MATH:OPTion:SENSitivity
Syntax
:MATH:OPTion:SENSitivity <sens>
:MATH:OPTion:SENSitivity?
Description
Set or query the sensitivity of the logic operation. The default unit is div (namely the
current vertical scale).
Parameter
Name
<sens>
Explanation
This command is only applicable to the logic operations (A&&B, A||B, A^B and !A).
Return Format
The query returns the sensitivity in scientific notation.
Example
:MATH:OPTion:SENSitivity 0.24
:MATH:OPTion:SENSitivity?
Related
:CHANnel<n>:SCALe
Commands
:MATH:OPERator
:MATH:OPTion:DIStance
Syntax
:MATH:OPTion:DIStance <dist>
:MATH:OPTion:DIStance?
Description
Set or query the smoothing window width of the differential operation (diff).
Parameter
Name
<dist>
Explanation
This command is only applicable to the differential operation (diff).
Return Format
The query returns an integer between 3 and 201.
Example
:MATH:OPTion:DIStance 25
Related
:MATH:OPERator
Command
:MATH:OPTion:ASCale
Syntax
:MATH:OPTion:ASCale <opt>
:MATH:OPTion:ASCale?
Description
Enable or disable the auto scale setting of the operation result or query the status of
the auto scale setting.
Parameter
Name
<opt>
Explanation
Return Format
The query returns 0 or 1.
Example
:MATH:OPTion:ASCale ON
2-44
Type
Range
0 to 0.96div, the step is 0.08div
Real
Type
Range
3 to 201
Integer
Type
Range
Bool
{{0|OFF}|{1|ON}}
When the auto scale is enabled, the instrument will automatically calculate the
vertical scale range according to the current operator, the vertical scale and the
horizontal timebase. If the current scale is out of the range, it will adjust the
vertical scale to the best value automatically.
Sending this command will modify the auto scale status of all the operations.
/*Set the sensitivity to 0.24div*/
/*The query returns 2.400000e-01*/
Chapter 2 Command System
Default
0
Default
3
Default
0|OFF
DS1000Z Programming Guide