RIGOL
:MEASure:ADISplay
Syntax
:MEASure:ADISplay <bool>
:MEASure:ADISplay?
Description
Enable or disable the all measurement function, or query the status of the all
measurement function.
Parameter
Name
<bool>
Explanation
The all measurement function can measure the following 20 parameters of the
source at the same time:
Voltage Parameters: Vmax, Vmin, Vpp, Vtop, Vbase, Vamp, Vavg, Vrms, Overshoot
and Peshoot
Time Parameters: Period, Frequency, Rise Time, Fall Time, + Width, - Width, + Duty
and - Duty
Other Parameters: Area and Period Area.
The all measurement function can measure CH1, CH2, CH3 and CH4 at the same
time. You can send the
measurement function.
Return
The query returns 0 or 1.
Format
Example
:MEASure:ADISplay ON
Related
:MEASure:AMSource
Command
:MEASure:AMSource
Syntax
:MEASure:AMSource <src>[,<src>[,<src>[,<src>]]]
:MEASure:AMSource?
Description
Set or query the source of the all measurement function.
Parameter
Name
<src>
Return
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.
Format
Example
:MEASure:AMSource CHANnel1,CHANnel3
2-56
Type
Range
Bool
{{0|OFF}|{1|ON}}
:MEASure:AMSource
Type
Range
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
Chapter 2 Command System
command to set the source of the all
DS1000Z Programming Guide
Default
0|OFF
Default
CHANnel1