:TRIGger:PULSe:LWIDth
Syntax
:TRIGger:PULSe:LWIDth <width>
:TRIGger:PULSe:LWIDth?
Description
Set or query the lower pulse width in pulse width trigger. The default unit is s.
Parameter
Name
<width>
Explanation
This command is available when the trigger condition (refer to :TRIGger:PULSe:WHEN) is
PGLess and NGLess.
Return
The query returns the lower pulse width in scientific notation.
Format
Example
:TRIGger:PULSe:LWIDth 0.000003
:TRIGger:PULSe:LWIDth?
Related
:TRIGger:PULSe:WHEN
Commands
:TRIGger:PULSe:UWIDth
:TRIGger:PULSe:LEVel
Syntax
:TRIGger:PULSe:LEVel <level>
:TRIGger:PULSe:LEVel?
Description
Set or query the trigger level in pulse width trigger. The unit is the same as the current
amplitude unit.
Parameter
Name
<level>
Explanation
For VerticalScale, refer to the
the
:CHANnel<n>:OFFSet
Return
The query returns the trigger level in scientific notation.
Format
Example
:TRIGger:PULSe:LEVel 0.16
:TRIGger:PULSe:LEVel?
Related
:CHANnel<n>:SCALe
Commands
:CHANnel<n>:OFFSet
2-90
Type
Range
Real
8ns to 9.99s
/*The query returns 3.000000e-06*/
Type
Range
(± 5 × VerticalScale from the screen center) -
Real
OFFSet
:CHANnel<n>:SCALe
command.
/*Set the trigger level to 160mV*/
/*The query returns 1.600000e-01*/
/*Set the lower pulse width to 3μs*/
command. For OFFSet, refer to
Chapter 2 Command System
Default
992ns
Default
0