RIGOL
:TRIGger:NEDGe:IDLE
Syntax
:TRIGger:NEDGe:IDLE <NR3>
:TRIGger:NEDGe:IDLE?
Description
Set or query the idle time in Nth edge trigger. The default unit is s.
Parameter
Return Format
The query returns the idle time in scientific notation.
Example
:TRIGger:NEDGe:IDLE 0.002
:TRIGger:NEDGe:IDLE?
:TRIGger:NEDGe:EDGE
Syntax
:TRIGger:NEDGe:EDGE <NR1>
:TRIGger:NEDGe:EDGE?
Description
Set or query the edge number in Nth edge trigger.
Parameter
Return Format
The query returns an integer between 1 and 65535.
Example
:TRIGger:NEDGe:EDGE 20
:TRIGger:NEDGe:LEVel
Syntax
:TRIGger:NEDGe:LEVel <level>
:TRIGger:NEDGe:LEVel?
Description
Set or query the trigger level in Nth edge trigger. The unit is the same as the current
amplitude unit.
Parameter
Explanation
For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
Return Format
The query returns the trigger level in scientific notation.
Example
:TRIGger:NEDGe:LEVel 0.16
:TRIGger:NEDGe:LEVel?
Related
:CHANnel<n>:SCALe
Commands
:CHANnel<n>:OFFSet
2-120
Name
Type
<NR3>
Real
Name
Type
<NR1>
Integer
Name
Type
<level>
Real
Range
16ns to 10s
/*Set the idle time to 2ms*/
/*The query returns 2.000000e-03*/
Range
1 to 65535
Range
(± 5 × VerticalScale from the screen center) -
OFFSet
/*Set the the trigger level to 160mV*/
/*The query returns 1.600000e-01*/
Chapter 2 Command System
Default
1μs
Default
2
Default
0