Rigol DS1000Z Series Programming Manual page 111

Digital oscilloscope
Hide thumbs Also See for DS1000Z Series:
Table of Contents

Advertisement

Chapter 2 Command System
:TRIGger:VIDeo:MODE
Syntax
:TRIGger:VIDeo:MODE <mode>
:TRIGger:VIDeo:MODE?
Description
Set or query the sync type in video trigger.
Parameter
Name
<mode>
Explanation
ODDField: trigger on the rising edge of the first ramp waveform pulse in the odd
field. This type is available when the video standard is NTSC or PAL/SECAM.
EVENfield: trigger on the rising edge of the first ramp waveform pulse in the even
field. This type is available when the video standard is NTSC or PAL/SECAM.
LINE: for the NTSC and PAL/SECAM video standards, trigger on the specified line in
the odd or even field; for the 480P and 576P video standards, trigger on the specified
line.
ALINes: trigger on all the horizontal sync pulses.
Return
The query returns ODDF, EVEN, LINE or ALIN.
Format
Example
:TRIGger:VIDeo:MODE ODDField
Related
:TRIGger:VIDeo:LINE
Commands
:TRIGger:VIDeo:STANdard
:TRIGger:VIDeo:LINE
Syntax
:TRIGger:VIDeo:LINE <line>
:TRIGger:VIDeo:LINE?
Description
Set or query the line number when the sync type in video trigger is LINE.
Parameter
Name
<line>
Explanation
Return
The query returns an integer.
Format
Example
:TRIGger:VIDeo:LINE 100
Related
:TRIGger:VIDeo:MODE
Command
DS1000Z Programming Guide
Type
Discrete
Type
Integer
NTSC: 1 to 525
PAL/SECAM: 1 to 625
480P: 1 to 525
576P: 1 to 625
Range
{ODDField|EVENfield|LINE|ALINes}
Range
Explanation
Refer to
RIGOL
Default
ALINes
Default
1